Bicon Dental Implants
Monday, September 8th 2008 4:30AM EDT

Orthodontic Implants and Components

Orthodontic Implants

Description Dia. Height Part Number
8.0mm Orthodontic Implant 2.5mm 8.0mm 260-325-208

 

 

Orthodontic Implant Components

Description Part Number
Orthodontic Implant Driver 260-101-907

 

 


Transitional Implant Considerations

Procedure:
• Drill a 2.0mm osteotomy to appropriate depth with Bicon’s 2.0mm pilot drill with external irrigation.
• Take care not to overheat the bone while drilling at ~1,100 RPM.
• Pilot bur is marked at 6.0mm, 8.0mm, 11.0mm and 14.0mm.*
• Be sure to drill to appropriate depth of intended implant to be placed or slightly beyond if the anatomy permits.
• Make note of the quality of bone.
• Check angulation or trajectory of osteotomy with a paralleling pin.
• Remove appropriate transitional implant from packaging and insert apical end of implant into osteotomy.
• Assemble appropriate transitional implant driver to Bicon’s 400:1 handpiece or a comparable slow speed latch-type handpiece.
• Slowly screw implant into the prepared osteotomy.
*Some pilot burs may have markings only at 8.0mm, 11mm, and 14mm. Check burs before drilling. (See Pilot Drill Reference).