Orthodontic Implants and Components
Orthodontic Implants
| Description |
Dia. |
Height |
Part Number |
| 8.0mm Orthodontic Implant |
2.5mm |
8.0mm |
260-325-208 |

Orthodontic Implant Components
| Description |
Part Number |
| Orthodontic Implant Driver |
260-101-907 |

Transitional Implant Considerations
Procedure:
- Drill a 2.0mm osteotomy to appropriate depth with Bicon’s 2.0mm pilot drill with external irrigation.
- Take care not to overheat the bone while drilling at ~1,100 RPM.
- Pilot bur is marked at 6.0mm, 8.0mm, 11.0mm and 14.0mm.*
- Be sure to drill to appropriate depth of intended implant to be placed or slightly beyond if the anatomy permits.
- Make note of the quality of bone.
- Check angulation or trajectory of osteotomy with a paralleling pin.
- Remove appropriate transitional implant from packaging and insert apical end of implant into osteotomy.
- Assemble appropriate transitional implant driver to Bicon’s 400:1 handpiece or a comparable slow speed latch-type handpiece.
- Slowly screw implant into the prepared osteotomy.
*Some pilot burs may have markings only at 8.0mm, 11.0mm, and 14.0mm. Check burs before drilling. If there is any doubt about the markings on any drill or reamer, take a measurement prior to using the reamer. (See Pilot Drill Reference).