Monday, September 6th 2010 11:31PM EDT

Orthodontic Implants and Components

Orthodontic Implants

Description Dia. Height Part Number
8.0mm Orthodontic Implant 2.5mm 8.0mm 260-325-208

 

Orthodontic Implant

 

Orthodontic Implant Components

Description Part Number
Orthodontic Implant Driver 260-101-907

 

Orthodontic Implant Driver

 

Transitional Implant Considerations

    Procedure:
  • Drill a 2.0mm osteotomy to appropriate depth with Bicon’s 2.0mm pilot drill with external irrigation.
  • Take care not to overheat the bone while drilling at ~1,100 RPM.
  • Pilot bur is marked at 6.0mm, 8.0mm, 11.0mm and 14.0mm.*
  • Be sure to drill to appropriate depth of intended implant to be placed or slightly beyond if the anatomy permits.
  • Make note of the quality of bone.
  • Check angulation or trajectory of osteotomy with a paralleling pin.
  • Remove appropriate transitional implant from packaging and insert apical end of implant into osteotomy.
  • Assemble appropriate transitional implant driver to Bicon’s 400:1 handpiece or a comparable slow speed latch-type handpiece.
  • Slowly screw implant into the prepared osteotomy.

    *Some pilot burs may have markings only at 8.0mm, 11.0mm, and 14.0mm. Check burs before drilling. If there is any doubt about the markings on any drill or reamer, take a measurement prior to using the reamer. (See Pilot Drill Reference).